Chapter 1: General Information
Recommended test equipment
1–16
Recommended test equipment
The following table is a list of the test equipment required to test performance, calibrate and
adjust, and troubleshoot this instrument. The table indicates the critical specification of the test
equipment and for which procedure the equipment is necessary. Equipment other than the
recommended model may be used if it satisfies the critical specification listed in the table.
Recommended Test Equipment
Equipment Required
Critical Specifications
Recommended Model
Use *
Signal Generator
1 - 2.25 GHz, sine wave, amplitude 30 - 200
mV
rms
, time base accuracy 0.25 ppm
Agilent 8664A
P
RF Amplifier
22 dB gain at 10 MHz, 1.3 GHz bandwidth
Agilent 8447D/F
P
Power Meter/Power
Sensor
1 - 500 MHz, -70 dBm to +44 dBm,
±
3% accuracy
Agilent EPM-
441A/Agilent 8482A
P
DMM
6 1/2 digit (0.1 mV) resolution, dcV accuracy
8 ppm/year, 4-wire resistance acc.
±
0.25%
Agilent 34401A
P, A, T
Power Supply
7 mV - 30 V dc, 0.1 mV accuracy and resolution
Agilent 6114A
P
Power Splitter
50
Ω
type N, outputs differ by <0.15 dB
Agilent 11667A
P
Probe
No substitute
Agilent 1161A
P
Probe Tip Adapter
1160 series to BNC
Agilent 5063-2143
P
Oscilloscope
General-purpose
Agilent 54622A
P, T
Blocking Capacitor
0.18
µ
F
Agilent 10240B
P
Cable
Type N (m)(m) - 3 foot
Agilent 11500A or B
P
Cable (2)
BNC - 3 foot
Agilent 10503A
P, A, T
Cable (3)
BNC - 9 inch
Agilent 10502A
P, A, T
Adapter
N (m) to BNC (m)
Agilent 1250-0082
P
Adapter
N (m) to BNC (f)
Agilent 1250-0780
P
Adapter
N (f) to BNC (m)
Agilent 1250-0077
A
Adapter (2)
BNC tee (m)(f)(f)
Agilent 1250-0781
P, T
Adapter
BNC (f)(f)
Agilent 1250-0080
T
Adapter (2)
BNC (f) to dual banana (m)
Agilent 1251-2277
P
Termination
BNC connectors 50
Ω
Agilent 10100C
P
Shorting cap
BNC
Agilent 1250-0774
P
Resistor (2)
5 Ω,
5 W
Agilent 0812-0047
T
Video monitor
Accepts VGA-standard video signals
T
Keyboard
PC-compatible, AT 5-pin connector
Agilent E2610A
T
Mouse
PS/2 compatible
Agilent C3751-60201
P, T
POST Card
Power-on self test card compatible with PC-
compatible systems
T
Fan Safety Shield
Clips onto side of chassis with cover removed
Agilent 54801-00601
A, T
Service software
No substitution
Agilent 54810-68700
T
Note: the Fan Safety Shield and Service software are part of the Service Kit, Agilent P/N 54810-68703.
* P = Performance Tests, A = Adjustments, T = Troubleshooting
Summary of Contents for 54835A
Page 8: ...Contents 4 ...
Page 64: ...3 26 ...
Page 78: ...4 12 ...
Page 138: ...5 60 ...
Page 186: ...Chapter 7 Replaceable Parts Exploded Views 7 14 Attenuator Assembly ...
Page 194: ...7 22 ...
Page 204: ...8 10 ...