3- 20
Agilent 54642A/42D Performance Test Record
Agilent 54642A/54642D
Serial No. ______________________________________
Test by _____________________________
Test Interval ____________________________________
Work Order No. ______________________
Recommended Next Testing ________________________
Temperature ____________
Threshold
Specification
Limits
Ch D7-D0
Ch D15-D8
Accuracy Test
5 V - 250 mV
4.750 V
________
________
(100 mV + 3% of
threshold setting)
5 V + 250 mV
5.250 V
________
________
-5 V - 250 mV
-5.250 V
________
________
-5 V + 250 mV
-4.750 V
________
________
0 V - 100 mV
-100 mV
________
________
0 V + 100 mV
100 mV
________
________
Voltage Measurement Accuracy
Range
Power Supply Setting
Test Limits
Channel 1
Channel 2
5 V/Div
35 V
34.04 V to 35.96 V
________
________
2 V/Div
14 V
13.616 V to 14.384 V
________
________
1 V/Div
7 V
6.808 V to 7.192 V
________
________
500 mV/Div
3.5 V
3.404 V to 3.596 V
________
________
200 mV/Div
1.4 V
1.3616 V to 1.4384 V
________
________
100 mV/Div
700 mV
680.8 mV to 719.2 mV
________
________
50 mV/Div
350 mV
340.4 mV to 359.6 mV
________
________
20 mV/Div
140 mV
136.16 mV to 143.84 mV
________
________
10 mV/Div
70 mV
68.08 mV to 71.92 mV
________
________
5 mV/Div
35 mV
34.04 mV to 35.96 mV
________
________
2 mV/Div
14 mV
13.232 mV to 14.768 mV
________
________
Bandwidth
Test Limits
Channel 1
Channel 2
3 dB at 500 MHz
________
________
Horizontal
∆
t and 1/
∆
t Accuracy
Generator Setting
Test Limits
Results
Frequency
10 kHz
9.98 kHz to 10.02 kHz
________
Period
100
µ
s
99.79
µ
s to 100.2
µ
s
________
Frequency
10 MHz
9.98 MHz to 10.02 MHz
________
Period
100 ns
99.77 ns to 100.2 ns
________
Frequency
500 MHz
492.6 MHz to 507.6 MHz ________
Period
2 ns
1.970 ns to 2.030 ns
________
Trigger Sensitivity
Test Limits
Channel 1
Channel 2
Internal trigger
25 MHz at 1 division
________
________
500 MHz at 0.6 divisions ________
________
Ext
External trigger
500 MHz at <200 mV
p-p
________
100 MHz at <100 mV
p-p
________
service.book Page 20 Wednesday, August 28, 2002 3:25 PM
Summary of Contents for 54641A
Page 4: ...iv service book Page iv Wednesday August 28 2002 3 25 PM ...
Page 7: ...1 General Information ...
Page 24: ...1 18 ...
Page 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 45: ...3 Testing Performance service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 66: ...3 22 service book Page 22 Wednesday August 28 2002 3 25 PM ...
Page 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 76: ...4 10 service book Page 10 Wednesday August 28 2002 3 25 PM ...
Page 77: ...5 Troubleshooting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 90: ...5 14 service book Page 14 Wednesday August 28 2002 3 25 PM ...
Page 91: ...6 Replacing Assemblies service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 107: ...7 Replaceable Parts service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 114: ...7 8 service book Page 8 Wednesday August 28 2002 3 25 PM ...