4- 7
Calibrating and Adjusting
To perform User Cal
To perform User Cal
Perform self-calibration:
• every 6 months or after 1000 hours of operation.
• if the ambient temperature is >10 °C from the calibration temperature.
• if you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience with other
instruments help determine if you need shorter adjustment intervals.
User Cal performs an internal self-alignment routine to optimize the signal path
in the oscilloscope. The routine uses internally generated signals to optimize
circuits that affect channel sensitivity, offset, and trigger parameters.
Disconnect all inputs and allow the oscilloscope to warm up before performing
this procedure.
User Cal should be performed at least once a year, any time the ambient
temperature of the oscilloscope has changed more than 10 °C since the last User
Cal, or after any repair.
Performing User Cal will not invalidate your Certificate of Calibration.
Successful completion of User Cal does not certify this oscilloscope with a
National Institute of Standards and Technology (NIST) calibration.
1
Disconnect all inputs signals from the oscilloscope
2
Set the rear-panel
CALIBRATION
switch to
UNPROTECTED
.
3
Press the
Utility
key, then press the
Service
softkey.
4
Begin the Self Cal by pressing the
User Cal
softkey.
5
When the User Cal is completed, set the rear-panel
CALIBRATION
switch
to
PROTECTED
.
service.book Page 7 Wednesday, December 18, 2002 8:35 AM
Summary of Contents for 54621A
Page 4: ...iv service book Page iv Wednesday December 18 2002 8 35 AM ...
Page 7: ...1 General Information service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 24: ...1 18 service book Page 18 Wednesday December 18 2002 8 35 AM ...
Page 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 44: ...2 20 service book Page 20 Wednesday December 18 2002 8 35 AM ...
Page 45: ...3 Testing Performance service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 66: ...3 22 service book Page 22 Wednesday December 18 2002 8 35 AM ...
Page 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 76: ...4 10 service book Page 10 Wednesday December 18 2002 8 35 AM ...
Page 77: ...5 Troubleshooting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 90: ...5 14 service book Page 14 Wednesday December 18 2002 8 35 AM ...
Page 91: ...6 Replacing Assemblies service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 107: ...7 Replaceable Parts service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 114: ...7 8 service book Page 8 Wednesday December 18 2002 8 35 AM ...