4- 4
Calibrating and Adjusting
To adjust the power supply
To adjust the power supply
The oscilloscope power supply has both a +3.3 V adjustment, and a
+5.1 V / -5.2 V balance adjustment. Other oscilloscope voltages are based on
the +3.3 V adjustment.
In this procedure, you will use a digital multimeter to measure the +3.3 V and
+5.1 V / -5.2 V test points, and adjust the power supplies to be within tolerance,
if necessary.
Table 4-1
Equipment Required to Adjust the Power Supply
1
Prepare the oscilloscope for the voltage adjustment.
a
Turn off the oscilloscope and disconnect the power cable.
b
Remove the oscilloscope cover. For a list of parts, see the “Replaceable
Parts” chapter, then return here.
c
Place the oscilloscope on its side.
d
Connect the negative lead of the digital multimeter to a ground point on the
oscilloscope chassis.
e
Reconnect the power cable.
f
Turn on the oscilloscope.
Equipment
Critical Specifications
Recommended Model/Part
Digital multimeter
0.1 mV resolution, accuracy ±0.05%
Agilent 34401A
service.book Page 4 Wednesday, December 18, 2002 8:35 AM
Summary of Contents for 54621A
Page 4: ...iv service book Page iv Wednesday December 18 2002 8 35 AM ...
Page 7: ...1 General Information service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 24: ...1 18 service book Page 18 Wednesday December 18 2002 8 35 AM ...
Page 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 44: ...2 20 service book Page 20 Wednesday December 18 2002 8 35 AM ...
Page 45: ...3 Testing Performance service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 66: ...3 22 service book Page 22 Wednesday December 18 2002 8 35 AM ...
Page 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 76: ...4 10 service book Page 10 Wednesday December 18 2002 8 35 AM ...
Page 77: ...5 Troubleshooting service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 90: ...5 14 service book Page 14 Wednesday December 18 2002 8 35 AM ...
Page 91: ...6 Replacing Assemblies service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 107: ...7 Replaceable Parts service book Page 1 Wednesday December 18 2002 8 35 AM ...
Page 114: ...7 8 service book Page 8 Wednesday December 18 2002 8 35 AM ...