30
53210A/53220A/53230A
Option
Installation
53210A/53220A/53230A Option Installation
7. Install the supplied SMA semi- rigid cable as follows:
•
Insert the end of the supplied semi- rigid cable with the female SMA
connector through the hole on the rear panel marked "Opt 106/115".
•
Line up the male SMA cable connector on the other end of the
semi- rigid cable and the Channel 2 assembly SMA edge connector,
connect them together and tighten the SMA cable connector using a
5/16" wrench.
•
Secure the rear panel SMA connector with the supplied 5/16" hex
nut and lockwasher. Do not overtighten.
8. Re- install the Option 300 Internal Battery (if moved). Be sure to
re- connect any disconnected cables.
9. Re - install the cover, rear bezel, and handle (if applicable).
10. Install the supplied Option 106 label above the rear panel SMA con-
nector in the location provided.
N O T E
Be sure there is no stress on the semi-rigid cable.