Technical Information
Serial
5
Agilent 4UHV Ion Pump Controller User Manual
/
87-900-137-01 (A)
85/104
Error code in decimal
format
Binary Description
32 0000000100000
Interlock
cable
64 0000001000000
Overtemp
HV
128 0000010000000
Protect
256 0000100000000
Measurement
error
512
0001000000000
HV out error
1024 0010000000000
Short
circuit
2048 0100000000000
HV
disabled
Example:
Interlock cable
Error code: 32 (decimal)
´
0000000010000 ( binary)
´
Interlock
cable error
Interlock cable
+ Overtemp HV
Error code: 96 (decimal)
´
0000000110000 ( binary)
´
Interlock
cable error + Overtemp HV
Error code (Win 206)
Error interlock
02 80
32 30 36 30 03 38 37
STX ADDR WINDOW RD ETX CRC
02 80
32 30 36 30 30 30 30 30 33 32 03 38 36
STX
ADDR
WINDOW
WR DATA ETX
CRC
Channel selection (Win 505)
Read
02 80
35 30 35 30 03 38 33
STX ADDR WINDOW RD ETX CRC
02 80
35 30 35 30 30 30 30 30 30 30 03 38 33
STX
ADDR
WINDOW
WR DATA ETX
CRC