Instructions for Use
Usage
50/104
Agilent 4UHV Ion Pump Controller User Manual
/
87-900-137-01 (A)
WARNING!
If an HV cable without interlock is to be used, the accessories bag contains a
specific kit of preassembled interlock leads. However, with this type of use,
the safety capability furnished using the interlock is lost.
Refer to the figure below for correct assembly of the interlock cable
furnished in the accessories bag in the case in which an HV cable
with interlock is not used.
Figure 3
NOTE
The cable interlock closes on pump ground. The high voltage is disabled if the
connection is interrupted.
The cable interlock closes on pump ground. The high voltage is
disabled if the connection is interrupted. To connect the controller to
the pump, use a HV cable with interlock (see parts that can be
ordered).
Correct connection of masses, of the HV cable between the controller
and pump and of the interlock cable is illustrated in the
“Connections” figure.