Chapter 3
21
Operation
4294A Setup
Figure 3-1
Extending Connector Sleeve
7. Connect the OPEN standard to the 7-mm OUTPUT port connector.
CAUTION
Never store the OPEN standard loosely in a box or drawer because this may result
in damage to the mating plane surfaces. Do not drop the OPEN standard or give it
any strong mechanical shock.
Figure 3-2
Connecting OPEN Standard
8. Press
PHASE COMP [-]
key to start the phase compensation data measurement.
When the phase compensation data measurement is completed, the softkey label