Figure
3-8.
Connecting
DUT
to
the
16034E
7.
P
erform
the
measurement.
Press
4
DISPLA
Y
F
ORMA
T
5 .
The
Internal
trigger
starts
and
continuously
repeats
the
measurement.
The
measured
L
s
and
R
s
values
of
the
ferrite-cored
inductor
are
displayed
in
large
characters
as
shown
in
Figure
3-9.
<MEAS
DISPLAY>
SYS
MENU
FUNC
:
Ls-Rs
RANGE:
AUTO
FREQ
:25.2000MHz
BIAS
:
0.000
V
LEVEL:*
10.0mA
INTEG:
LONG
L
s
:
4
5
5
.
6
8
7
n
H
R
s
:
1
.
6
6
4
1
9
Vm
:
957.5mV
Im
:
10.01mA
CORR:OPEN,SHORT
CBL:0m
Figure
3-9.
Measurement
Results
of
The
F
errite-Cored
Inductor
Basic
Measurement
Examples
3-9