DC
V
oltage
Biased
Measurement
When
the
4285A
is
equipped
with
Option
001
DC
Bias
,
640V
(100
mA)
DC
bias
can
be
applied
to
the
DUT
.
In
this
example
,
a
470
pF
capacitor
will
be
measured
under
the
following
conditions
.
DUT
:
470
pF
ceramic
capacitor
(Agilent
P
art
Number
0160-4808)
Measurement
Conditions
Function
:
C
p
-
D
Frequency
:
5
MHz
T
est
Signal
Level
:
1
V
(Constant)
DC
Bias
Level
:
5
V
Equipment
Required
4285A
with
Option
001
16047C
T
est
Fixture
Shorting
bar
(Agilent
P
art
Number
5000-4226)
or
equivalent
Preparation
Before
Measurement
Checking
the
Option
Installed
First
of
all,
conrm
that
your
4285A
is
equipped
with
the
Option
001.
T
o
do
this
,
press
4
CA
T
ALOG/SYSTEM
5
and
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
CONFIG
,
and
check
the
DC
BIAS
(#001)
monitor
area.
When
INSTALLED
is
displayed
in
this
area,
your
4285A
is
equipped
with
Option
001.
Operation
Procedure
1.
Turn
the
4285A
ON.
2.
Set
the
4285A
to
the
normal
LCR
measurement
mode
which
uses
the
auto
balancing
bridge
measuring
method.
a.
Press
4
CA
T
ALOG/SYSTEM
5
and
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
CONFIG
.
SY
STEM
CONFIG
page
will
now
be
displayed
b.
Move
the
CURSOR
to
the
CONFIG
eld.
NNNNNNNNNNNNNNNNNNNN
I
BIAS
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Q
ADAPTER
,
and
N
NNNNNNNNNN
OFF
will
be
displayed.
c.
Press
NNNNNNNNNNN
OFF
to
perform
the
normal
LCR
measurement.
3.
Setup
the
4285A
measurement
conditions
by
modifying
the
elds
on
the
MEAS
DISPLA
Y
page
.
Set
FREQuency
to
5
MHz
and
LEVEL
to
a
constant
1
V
.
(The
other
functions
will
be
left
at
the
default
settings
.)
a.
Move
the
CURSOR
to
the
FREQ
eld.
The
current
measurement
frequency
,
1.0000MHz ,
will
be
displayed
in
this
eld.
b.
Press
4
5
5 .
5
will
be
displayed
on
the
system
message
line
,
and
the
softkey
lables
are
changed
to
the
available
units
(
NNNNNNNNNNN
kHz
and
NNNNNNNNNNN
MHz
).
Press
NNNNNNNNNNN
MHz
.
5.0000MHz
will
be
new
displayed
in
the
FREQ
eld.
c.
Press
4
MEAS
SETUP
5 .
The
MEAS
SETUP
page
will
be
displayed.
3-10
Basic
Measurement
Examples