Introduction to Programming
4
Agilent 34945A Microwave Switch/Attenuator Driver Module User’s Guide
75
Figure 4-11. Y1151A Distribution Board Web User Interface Preset State.
ROUT:CHAN:LAB “SW 1 Path 1”,(@1101)
ROUT:CHAN:DRIV:PAIR:MODE OFF,(@1101)
ROUT:CHAN:VER:ENAB OFF,(@1101)
ROUT:CHAN:VER:POL NORM,(@1101)
ROUT:CHAN:DRIV:TIME:REC 0.0,(@1101)
ROUT:CHAN:DRIV:TIME:SETT 0.0,(@1101)
ROUT:CHAN:DRIV:PULS:MODE ON,(@1101)
ROUT:CHAN:DRIV:PULS:WIDTH 0.015,(@1101)
ROUT:CHAN:DRIV:OPEN:DEF (@1101)
ROUT:RMOD:BANK:LED:DRIV:ENAB ON,1,(@1100)
ROUT:RMOD:BANK:LED:DRIV:LEV 0.0050,1,(@1100)
ROUT:RMOD:BANK:DRIV:MODE OCOL,1,(@1100)
ROUT:RMOD:DRIV:SOUR OFF,(@1100)
ROUT:RMOD:DRIV:SOUR:BOOT OFF,(@1100)
ROUT:RMOD:DRIV:LIM 1,(@1100)
ROUT:RMOD:BANK:PRES 1,(@1100)
bank diagram
(right-click)
see Chapter 5 for the Y1151A
switch control table
(all channels in the bank are preset)