34941A and 34942A SCPI Programming Examples

Agilent 34941A-34942A RF Multiplexer Modules User’s Guide

3

34941A and 34942A SCPI Programming Examples

The programming examples below provide you with SCPI command 
examples to use for actions specific to the RF MUX switch modules. 

The slot and channel addressing scheme used in these examples follow the 
form 

sccc 

where 

s

 is the mainframe slot number (1 through 8) and 

ccc

 is 

the channel number. For information on specific configurations, refer to 
the simplified schematic on 

page 4

.

For complete information on the SCPI commands used to program the 
34980A, refer to the 

Agilent 34980A Programmer’s Reference

 contained on 

the 

34980A Product Reference

 CD. For example programs, also refer to the 

34980A Product Reference

 CD. 

Example: Closing channels  

You can only 

close

 channels on the RF MUX 

modules. When you close a channel, any already- closed channels open 
automatically. The relays switch in sequence to avoid momentary 
connection of the wrong input to the multiplexer output. The following 
command closes channel 03 on  Bank  1  of  an  RF  MUX  module  in  slot  5. 

ROUTe:CLOSe (@5103)

Example: Querying channels for open or close state 

The following commands 

returns the close or open state of channel 33 of a module in slot 5. 

ROUT:CLOSe? (@5033)

ROUT:OPEN? (@5033)

Example: Querying the system for module identify 

The following command 

returns the identify of the module installed in slot 7. 

SYSTem:CTYPe? 7

Example: Reading the cycle count for a relay 

On these modules, each bank 

consists of two 

leaf

 relays and one 

tree

 relay (see the simplified schematic 

on 

page 4

). The module stores the cycle count for each of the three relays 

on all four banks. The cycle count is the greater of the three values on the 
specified bank (i.e., reflecting the cycle count for the entire bank). 
Therefore, the count for Channels 101, 102, 103, and 104 will 

always

 be 

equal. The following statement reads back the number of completed cycles 
for the channels 101 and 202 on a module installed in slot 6. 

DIAGnostic:RELay:CYCLes? (@6101,6202)

Example: Clearing the cycle count for a relay  

The following command resets 

the cycle count on the channels 103 and 201 for a module in slot 1. 

Note that clearing the cycle count on a specific channel will clear the 
count on 

all

 three relays in the corresponding bank.

DIAGnostic:RELay:CYCLes:CLEar (@1103,1201)

Example: Resetting module to power-on state  

The following command resets a 

module  in  slot  4  to  its  power- on  state. 

SYSTem:CPON 4

Summary of Contents for 34941A

Page 1: ...and service on the web www keysight com Select your country from the drop down menu at the top Under Electronic Test and Measurement click on Services The web page that appears next has contact infor...

Page 2: ......

Page 3: ...Agilent Technologies Agilent 34941A 34942A RF Multiplexer Modules User s Guide Agilent Technologies Inc Printed in Malaysia Edition 2 September 2012 E0912 34980 90041 34980 90041...

Page 4: ......

Page 5: ...nformation contained herein Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms i...

Page 6: ...uits before removing the instrument cover Do Not Modify the Instrument Do not install substitute parts or perform any unauthorized modification to the prod uct Return the product to an Agilent Sales a...

Page 7: ...Internet There are two methods to obtain the Declaration of Conformity Go to http regulations corporate agilent com DoC search htm You can then search by product number to find the latest Declaration...

Page 8: ...34941A 34942A RF Multiplexer Modules User s Guide v Contents RF Multiplexer Switch Modules 1 Operating Considerations 2 34941A and 34942A SCPI Programming Examples 3 34941A and 34942A Simplified Sche...

Page 9: ...vi Agilent 34941A 34942A RF Multiplexer Modules User s Guide...

Page 10: ...between the two RF MUX modules lie in their characteristic impedance and their use of connectors external cables are not provided with the module The 34941A the 50 version uses SMA connectors When ins...

Page 11: ...UX modules you automatically close all relays that create a direct path to the Common of a bank Channel Relay Operation With RF MUX switches you cannot open switches programmatically You can only clos...

Page 12: ...channels for open or close state The following commands returns the close or open state of channel 33 of a module in slot 5 ROUT CLOSe 5033 ROUT OPEN 5033 Example Querying the system for module ident...

Page 13: ...h contains four banks of latching switches Each bank consists of three Form C relays The front panel of the two RF MUX modules are similar with channel labels in the same positions the unique product...

Page 14: ...C channel relays 2 connectors 1 D description 1 E electrical isolation 2 G grounding 2 M mini SMB connectors 1 module descriptions 1 module impedance 34941A 1 34942A 1 P programming examples 3 R rela...

Page 15: ...6 Agilent 34941A 34942A RF Multiplexer Modules User s Guide Index...

Reviews: