Agilent Technologies 34941A User Manual Download Page 11

2

Agilent 34941A-34942A RF Multiplexer Modules User’s Guide

Operating Considerations.

Operating Considerations.

Electrical Considerations

See the 

Introduction to the Plug In Modules

 chapter of the 34980A 

Mainframe User’s Guide

 

for detailed environmental operating conditions 

for the 34980A mainframe and its installed modules. That guidance sets 
maximum per channel current and power ratings at rated voltage for 
pollution degree 1 (dry) and pollution degree 2 (possible condensation) 
conditions, for each of the GP modules.

Signal Connections

The RF MUX modules do not connect to the analog buses. Instead, all 
signal connections are made through the visible connectors via external 
cables. Each visible connector on an RF MUX module is labeled with a 
number (11 through 44), that represents a channel you can close 
programmatically from the front panel or with using the Web Browser 
Interface. When you close a channel on the RF MUX modules you 
automatically close all relays that create a direct path to the Common of a 
bank.

Channel Relay Operation

With RF MUX switches, you cannot open switches programmatically. You 
can only close a channel. When you close one channel, another channel 
automatically opens. Therefore, only one channel relay in each bank is 
closed at any time. 

Electrical Isolation

You can configure each bank on the RF MUX modules to be either isolated 
or chassis- grounded. The modules come with chassis- grounded metal 
shoulder washers installed on all connectors in each bank of relays. If you 
want to isolate a bank from the other banks and from chassis- ground, you 
must remove the five metal washers in that bank and replace them with 
the provided plastic shoulder washers. 

To isolate a bank from other banks 
and chassis-ground, install the 
provided 

plastic

 shoulder washers 

on the connectors in the bank to 
be isolated. When installing the 
plastic shoulder washers, use 

7 in-lbs of torque. 

The 34941A and 34942A are shipped 
from the factory chassis-grounded 

with 

metal

 shoulder washers 

installed on all connectors in each 
bank of relays. 

Summary of Contents for 34941A

Page 1: ...and service on the web www keysight com Select your country from the drop down menu at the top Under Electronic Test and Measurement click on Services The web page that appears next has contact infor...

Page 2: ......

Page 3: ...Agilent Technologies Agilent 34941A 34942A RF Multiplexer Modules User s Guide Agilent Technologies Inc Printed in Malaysia Edition 2 September 2012 E0912 34980 90041 34980 90041...

Page 4: ......

Page 5: ...nformation contained herein Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms i...

Page 6: ...uits before removing the instrument cover Do Not Modify the Instrument Do not install substitute parts or perform any unauthorized modification to the prod uct Return the product to an Agilent Sales a...

Page 7: ...Internet There are two methods to obtain the Declaration of Conformity Go to http regulations corporate agilent com DoC search htm You can then search by product number to find the latest Declaration...

Page 8: ...34941A 34942A RF Multiplexer Modules User s Guide v Contents RF Multiplexer Switch Modules 1 Operating Considerations 2 34941A and 34942A SCPI Programming Examples 3 34941A and 34942A Simplified Sche...

Page 9: ...vi Agilent 34941A 34942A RF Multiplexer Modules User s Guide...

Page 10: ...between the two RF MUX modules lie in their characteristic impedance and their use of connectors external cables are not provided with the module The 34941A the 50 version uses SMA connectors When ins...

Page 11: ...UX modules you automatically close all relays that create a direct path to the Common of a bank Channel Relay Operation With RF MUX switches you cannot open switches programmatically You can only clos...

Page 12: ...channels for open or close state The following commands returns the close or open state of channel 33 of a module in slot 5 ROUT CLOSe 5033 ROUT OPEN 5033 Example Querying the system for module ident...

Page 13: ...h contains four banks of latching switches Each bank consists of three Form C relays The front panel of the two RF MUX modules are similar with channel labels in the same positions the unique product...

Page 14: ...C channel relays 2 connectors 1 D description 1 E electrical isolation 2 G grounding 2 M mini SMB connectors 1 module descriptions 1 module impedance 34941A 1 34942A 1 P programming examples 3 R rela...

Page 15: ...6 Agilent 34941A 34942A RF Multiplexer Modules User s Guide Index...

Reviews: