BM:INTernal:RATE {<
frequency
>|MINimum|MAXimum}
Set the burst rate for internally triggered bursts. The burst rate
frequency defines the interval between bursts. Select from 10 mHz to
50 kHz.
The default is 100 Hz
.
MIN
= 10 mHz.
MAX
= 50 kHz.
[
Stored in volatile memory
]
The burst rate setting is used only when
internal
triggering is
enabled (triggered mode). The burst rate is ignored when single
triggering or external triggering is enabled.
It is possible to specify a burst rate which is too fast for the function
generator to output with the specified carrier frequency and burst
count. If the burst rate is too high, the function generator will
internally adjust it as needed to continuously re-trigger the burst.
The adjustment is handled internally by the function generator
(the burst rate displayed or queried will be the same as specified).
BM:INTernal:RATE? [MINimum|MAXimum]
Query the burst rate. Returns a value in hertz.
BM:SOURce {INTernal|EXTernal}
Select the burst modulation source. In the
external gated burst
mode,
the output waveform is either “on” or “off ” based on the level of the
external signal applied to the rear-panel
Ext Trig
terminal.
The default
is INT
. [
Stored in volatile memory
]
When the
internal
burst source is selected, the external gated mode
is disabled.
When the
external gate
source is selected, the output is enabled or
disabled based on the logic level of the gate signal applied to the
Ext Trig
terminal. When the gate signal is true (
TTL
high),
the function generator outputs a continuous waveform. When the
gate signal is false (
TTL
low), the output is turned off (zero volts or
the dc offset level).
When the
external gate
source is selected, the burst count, burst rate,
burst phase, and burst trigger source are ignored (these parameters
are used for the triggered burst mode only).
BM:SOURce?
Query the present burst modulation source. Returns “
INT
” or “
EXT
”.
Chapter 4 Remote Interface Reference
Burst Modulation Commands
164
Summary of Contents for 33120A
Page 2: ......
Page 10: ...8 ...
Page 14: ...Contents Contents 12 ...
Page 15: ...1 Quick Start 1 ...
Page 31: ...2 Front Panel Menu Operation 2 ...
Page 55: ...3 Features and Functions 3 ...
Page 126: ...124 ...
Page 127: ...4 Remote Interface Reference 4 ...
Page 204: ...SCPI Status System Chapter 4 Remote Interface Reference The SCPI Status Registers 202 ...
Page 228: ...226 ...
Page 229: ...5 Error Messages 5 ...
Page 244: ...242 ...
Page 245: ...6 Application Programs 6 ...
Page 272: ...270 ...
Page 273: ...7 Tutorial 7 ...
Page 299: ...8 Specifications 8 ...
Page 304: ...302 ...