Chapter 1 Quick Start
If the Power Supply Does Not Turn On
20
If the Power Supply Does Not Turn On
Use the following steps to help solve problems you might encounter when
turning on the instrument. If you need more help, refer to chapter 5 for
instructions on returning the instrument to Agilent Technologies for service.
1 Verify that there is ac power to the power supply.
First, verify that the power cord is firmly plugged into the power receptacle on
the rear panel of the power supply. You should also make sure that the power
source you plugged the power supply into is energized. Then, verify that the
power supply is turned on.
2 Verify the power-line voltage setting.
The line voltage is set to the proper value for your country when the power
supply is shipped from the factory. Change the voltage setting if it’s not correct.
The settings are: 100, 115, or 230 Vac.
3 Verify that the correct power-line fuse is installed.
The correct fuse is installed for your country when the power supply is shipped
from the factory. See the table below to replace the fuse for your power supply.
See the next page if you need to change the line voltage setting and the power-
line fuse.
Model
Agilent Part Number
Part Description
E3646A/47A
2110-1070
Fuse 2.5A T 125V for 100 and 115 Vac
E3646A/47A
2110-0457
Fuse 1A T 250V for 230Vac
E3648A/49A
2110-0996
Fuse 4A T 250V for 100 and 115 Vac
E3648A/49A
2110-1068
Fuse 2A T 250V for 230 Vac
Summary of Contents for 2110-0457
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 209: ...208 Index Index...