Chapter 4 Remote Interface Reference
The SCPI Status Registers
99
4
Using such a status register configuration allows a status event to be cross-
referenced by output and type of event. The
INSTrument
register indicates
which output(s) have generated an event. The
ISUMmary
register is a pseudo-
questionable status register for a particular logical output.
+
STATus:QUEStionable:INSTrument:ISUMmary1
Event Registers Enable Registers
Voltage
Current
0
1
"OR"
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Overvoltage
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
9
+
STATus:QUEStionable:INSTrument:ISUMmary2
Event Registers Enable Registers
Voltage
Current
0
1
"OR"
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Overvoltage
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
9
+
STATus:QUEStionable:INSTrument
Event Registers Enable Registers
INST1 Event Summary
2
1
"OR"
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
INST2 Event Summary
Not Used
To STATus:QUEStionable, bit 13
Summary of Contents for 2110-0457
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 209: ...208 Index Index...