Manual Change
Source Voltage Accuracy Test
Figure 6
Resistance Measurement Accuracy Test Setup (Floating)
Step 5.
Set the RC Box Adapter switch to the
F
position.
Step 6.
Press [blue][5]. Select
ExecCal
and press [Enter] to perform the calibration.
Step 7.
Set the source voltage to 100V using the following procedure:
1. Press [Source Voltage] to display the source voltage setup menu.
2. Press [1] [0] [0] [Enter] to set the source voltage to 100V.
3. Confirm that + 100V is displayed on the right side of the LCD.
Step 8.
Press [V Output] to turn the source voltage ON.(V Output indicator turns ON.)
Step 9.
Press [blue] [4]. Select
OpenMeas
and press [Enter] to perform the OPEN correction.
Step 10.
Press [V Output] to turn the source voltage OFF.(V Output indicator turns OFF.)
Step 11.
Connect the 16353C to the RC Box Adapter as shown in Figure 7.
Figure 7
16353x Connection (Floating)
4339boe1007
4339B
: No Connection
RC Box
Adapter
4339boe1004
RC Box Adapter
Direct connection, no cable
BNC Cable
16353x