![Agilent Technologies 16353A Manual Change Download Page 13](http://html1.mh-extra.com/html/agilent-technologies/16353a/16353a_manual-change_2867821013.webp)
Manual Change
Source Voltage Accuracy Test
Table 3
Current Measurement Accuracy Test Settings
Test Current
Standard
Resistor
Voltage
Setting
Measurement
Time
100pA
16353G
1V
LONG
1nA
16353F
1V
LONG
10nA
16353E
1V
LONG
100nA
16353D
1V
LONG
1
μ
A
16353C
1V
LONG
10
μ
A
16353B
1V
LONG
100
μ
A
16353B
10V
SHORT