Source Voltage Accuracy Test
Resistance Measurement Accuracy Test
The 4339B measures the resistance values of the calibrated 16353A/16353H, and these
measurements are compared with the 16353A/16353H’s calibration value.
Specification
Test Equipment
Procedure
WARNING
When the High VOltage indicator is lit, the 4339B outputs high voltages of up to
1000Vdc maximum.
Resistance Measurement Accuracy test (Floating Device).
Step 1.
Record the 16353A/16353H Standard Resistor Set calibration values into the Calculation
Sheet.
Step 2.
Reset the 4339B using the following procedure:
1. Press [blue] [
⋅
] to display the system reset menu.
2. Select Yes using [
←
][
↓
] or [
↑
][
→
] and press [Enter].
Step 3.
Set the Offset-error Canceling to ON using the following procedure:
1. Press [blue][-] to display the configuration setting menu.
2. Select Offset using [
←
][
↓
] or [
↑
][
→
] and press [Enter].
3. Select On using [
←
][
↓
] or [
↑
][
→
] and press [Enter].
4. Select Exit and press [Enter] to exit the configuration setting menu.
Step 4.
Set up the equipment as shown in Figure 6.
Basic Measurement Accuracy:
±
0.6% (See Chapter 8 General Information for
details.)
Description
Recommended model
Standard Resistor Set
16353A, 16353H
RC Box Adapter
PN 04339-65005