8-20 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver
Troubleshooting the A6 Interface Assembly
Troubleshooting the A6 Interface Assembly
The A6 interface assembly contains the GPIB address switch, the
connectors for the X and Y attenuators, and the wiring for switch
outputs 9 and 0.
To Troubleshoot the A6 Interface Assembly
1. To test the circuitry of the A6 interface assembly, check the driver
both in manual and remote operation.
a. Manually press the attenuator buttons several times. The light
should come on/off each time that a button is activated.
b. If the manual verification is satisfactory, check the
computer-controlled remote operation by sending these
commands.
2. If the response is satisfactory, the A6 interface assembly should be
operational.
3. If the response is unsatisfactory, refer to the troubleshooting
procedures for the A2 driver assembly.
Command
Desired Outcome
OUTPUT 728;”Al2345678”
All attenuator lights go on except the local light.
OUTPUT 728;”Bl2345678”
All attenuator lights go off except the remote light.