11713A Operating and Service Manual 4-7
Verification
GPIB Interface Check for Remote Operation
Local Lockout and Clear Lockout/ Set Local Messages
This check requires the bus to be in the remote state. Check that the
driver’s REMOTE light is ON.
Operator’s Response
Press the LOCAL key on the driver front panel. The REMOTE LED
should remain ON. If not, the driver failed to process the Local Lockout
message.
Operator’s Response
Check that the LOCAL light is ON and REMOTE light is OFF. Press
numbered pushbutton switches and check that switches change status
(LEDs on and off). If not, the driver failed to process the Clear
Lockout/Set Local message.
Description
Command
Local Lockout: Send Local Lockout message.
LOCAL LOCKOUT 7
Description
Command
Clear Lockout/Set Local:
LOCAL 7
Send Clear Lockout/Set Local message to clear
Local Lockout messages and set all devices to
Local