X, P, and K 11644A
2-5
Specifications
Electrical Specifications
Electrical Specifications
list the electrical specifications of the terminations, ¼
wavelength shims, and standard sections in your kit.
Measurement Uncertainty
Vector Network Analyzer (VNA) measurement accuracy depends on the accuracy of the
calibration standards, the calibration method employed, instrumentation accuracy and
stability, environmental factors, and the actual characteristic of the device under test
(DUT). Because of the complexity in these relationships, Agilent created uncertainty
computation tools to calculate the overall measurement uncertainty of a VNA system. In
the past, when the Agilent 8510 VNA system was being produced, a “Specification and
Verification” RMB program was included to perform an uncertainty calculation. Currently,
for the Agilent PNA and ENA families of network analyzers, a customer can download the
Table 2-4 Electrical Specifications of Terminations
Device
Frequency
Termination Specification
X11644A WR-90
8.2 to 12.4 GHz
return loss
≥
42 dB
P11644A WR-62
12.4 to 18 GHz
return loss
≥
42 dB
K11644A WR-42
18 to 26.5 GHz
return loss
≥
42 dB
Table 2-5 Electrical Specifications of
¼
Wavelength Shims
Part Number
Waveguide
Band
Return Loss
(dB)
a
a. Return loss specification is
guaranteed by mechanical
specifications.
11664-20021
WR90
55
11664-20020
WR62
11664-20019
WR42
Table 2-6 Electrical Specifications of Standard Sections
Device
Frequency
Standard Sections
Specification
X11644A WR-90
8.2 to 12.4 GHz
return loss
≥
42 dB
P11644A WR-62
12.4 to 18 GHz
return loss
≥
42 dB
K11644A WR-42
18 to 26.5 GHz
return loss
≥
42 dB
Summary of Contents for 11644A Series
Page 5: ...X P and K 11644A 1 1 1 General Information ...
Page 10: ...1 6 X P and K 11644A General Information Preventive Maintenance ...
Page 11: ...X P and K 11644A 2 1 2 Specifications ...
Page 17: ...X P and K 11644A 3 1 3 User Information ...
Page 24: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition ...
Page 25: ...X P and K 11644A 4 1 4 Use Maintenance and Care of the Devices ...
Page 34: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage ...
Page 35: ...X P and K 11644A 5 1 5 Performance Verification ...
Page 50: ...5 16 X P and K 11644A Performance Verification System Operation Checks ...
Page 51: ...X P and K 11644A 6 1 6 Troubleshooting ...
Page 55: ...X P and K 11644A 7 1 7 Replaceable Parts ...
Page 56: ...7 2 X P and K 11644A Replaceable Parts ...