X, P, and K 11644A
4-5
Use, Maintenance, and Care of the Devices
Connections
Connections
Good connections require a skilled operator. Slight errors in operator technique can have a
significant effect on measurements and measurement uncertainties. The most common
cause of measurement error is poor connections.
The following procedures illustrate how to make good connections.
Waveguide Devices
IMPORTANT
Unlike threaded devices, the WR-90, WR-62, WR-42 waveguide mating
planes are flanges (often precision) that you must carefully screw together.
Always connect waveguide in the same flange orientation. For example, use
the label as a reference and always connect a device with the label facing the
same direction.
Precision Flanges
A precision flange has four corner screw holes and two precision alignment holes, as shown
in
. A non-precision flange has only four corner screw holes.
Figure 4-2 Precision Alignment Holes
Summary of Contents for 11644A Series
Page 5: ...X P and K 11644A 1 1 1 General Information ...
Page 10: ...1 6 X P and K 11644A General Information Preventive Maintenance ...
Page 11: ...X P and K 11644A 2 1 2 Specifications ...
Page 17: ...X P and K 11644A 3 1 3 User Information ...
Page 24: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition ...
Page 25: ...X P and K 11644A 4 1 4 Use Maintenance and Care of the Devices ...
Page 34: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage ...
Page 35: ...X P and K 11644A 5 1 5 Performance Verification ...
Page 50: ...5 16 X P and K 11644A Performance Verification System Operation Checks ...
Page 51: ...X P and K 11644A 6 1 6 Troubleshooting ...
Page 55: ...X P and K 11644A 7 1 7 Replaceable Parts ...
Page 56: ...7 2 X P and K 11644A Replaceable Parts ...