Performance Verification
16
Figure 6
Equipement Setup
If the Probe Fails a Test
1
Replace the battery. Verify that the battery voltage is 9V ±0.1V.
2
Ensure that:
■
The probe's magnetic contact surfaces are clean
■
The probe tip closes completely around the coil's conductors
■
The coil is centered in the probe tip opening
■
The probe tip is perpendicular to the coil's conductors
■
The probe current flow indicator points in the correct direction
3
Demagnetize the probe by opening and closing the clamp 5 to 10
times.