Specifications
All probe specifications are warranted based on the following
conditions:
■
Within one year of calibration
■
23
°
C ±5°
C
■
20% to 75% relative humitidy
■
Probe zeroed
■
1 minute warm up
■
Battery at 9V ± 0.1V
■
External magnetic field < 40 A/m
■
No adjacent current carrying conductor
■
1 M
Ω
/100 pF probe termination
Characteristics
All probe characteristics are the typical performance values and are
not warranted. Characteristics are based on these conditions:
■
Within one year of calibration
■
23
°
C ±5°
C
■
20% to 75% relative humitidy
■
Probe zeroed
■
1 minute warm up
■
Battery at 9V ± 0.1V
Table 1
Specifications
Item
Specification
Measurement Accuracy
(Input Current: dc to 1 kHz)
100 mV/A
Sensitivity Setting
10 mV/A
Sensitivity Setting
50 mA to 10A
< ±3% of reading ±50 mA
—
50 mA to 40A
n/a
< ±4% of reading ±50 mA
40A to 100A
n/a
< ±15% of reading ±100 mA
Bandwidth
–3 dB
dc to 100 kHz
(with derating, see
“Typical Response Curves”
on page 13)