Service
1
Agilent 1000A Series Oscilloscopes Service Guide
33
1
Connect the equipment (see
Figure 10
).
a
Connect the signal generator output to the oscilloscope channel 1
input.
2
Verify the trigger sensitivity when the vertical scale is
≥
5 mV/div.
For each of the bandwidth boundaries (maximum bandwidth and
10 MHz):
a
Press the
[Default Setup]
key.
b
Set the output frequency of the signal generator to the bandwidth
boundary and set the amplitude to about 10 mVpp.
c
Press the
[Auto-Scale]
key.
d
Set the horizontal scale so that a few cycles of the waveform are
displayed.
e
Set the input channel’s probe attenuation to 1X.
f
Set the input channel’s vertical scale to 5 mV/div.
g
Decrease the amplitude from the signal generator to the specified
vertical division (1.5 div at maximum bandwith or 1 div at 10 MHz).
Figure 10
Connecting Equipment for Trigger Sensitivity Test
Oscilloscope
Signal Generator
BNC cable
Type-N (m) to
BNC (f) adapter
50
W
feed
through