Automatic exposure detection
The DR detector detects X-ray exposure to automatically perform the image
acquisition.
Before performing the exposure, the DR detector must be ready. Check the
status of the DR detector in the DR Detector Switch.
WARNING:
Do not hit or drop the equipment. If it receives a strong jolt,
image acquisition can be triggered without X-ray exposure.
WARNING:
Very short exposure time can cause failure to trigger the image
acquisition. Use an exposure time of at least 5 ms.
WARNING:
Applying collimation to leave only a very small area exposed can
cause failure to trigger the image acquisition.
WARNING:
Very low dose can cause failure to trigger the image acquisition.
A dose of at least 5 nGy is required.
WARNING:
Specific exposure conditions (use of grid, thickness of the
exposed object) can cause failure to trigger the image
acquisition or horizontal artifacts in the acquired image.