4. ERROR MESSAGES
During the routine use of the equipment, the display shows various messages about the status of the table.
This includes movement restrictions due to either room configuration or clearance space parameters or
present table status. Please, refer to the tables below.
4.1. INFORMATION MESSAGES
These messages appear on the top of touch screen of the main
table control console
Messages displayed on top of screen
description
What to do
MANUAL POSITIONING
This indicates that you are working in manual positioning mode instead of
automatic one
These messages are
informative and do not
require pressing any key.
These indications stay
displayed until any
change (other control ...).
If the message persists
when you order another
movement, turn the
system OFF and ON.
If the equipment remains
inoperative, turn it OFF
and call Field Service
POSITIONING IN PROGRESS
This indicates that the system is moving to reach the selected position
TABLE TOP MIN
This indicates that the tabletop reaches its end longitudinal position on foot side
TABLE TOP MAX
This indicates that the tabletop reaches its end longitudinal position on head side
TILTING LEFT LIMIT
This indicates that the table reaches its end tilting position in tredelenburg
TILTING RIGHT LIMIT
This indicates that the table reaches its end tilting position at +90°
S.I.D. LOWER LIMIT
This indicates that the tube reaches its lower position for minimum SID
S.I.D. UPPER LIMIT
This indicates that the tube reaches its higher position for maximum SID
CROSS COLLIMATOR MIN LIMIT
This indicates that the lateral shutters of collimator are closed completely
CROSS COLLIMATOR MAX LIMIT
This indicates that the lateral shutters of collimator are opened completely
LONG COLLIMATOR MIN. LIMIT
This indicates that the longitudinal shutters of collimator are closed completely
LONG COLLIMATOR MAX LIMIT Y
This indicates that the longitudinal shutters of collimator are opened completely
CARRIAGE MIN LIMIT
This indicates that the carriage reaches its end position on foot side
CARRIAGE MAX LIMIT
This indicates that the carriage reaches its end position on head side
COLUMN ANGULATION MIN LIMIT
This indicates that the column reaches its end angulation position at -40°
COLUMN ANGULATION MAX LIMIT This indicates that the column reaches its end angulation position at +40°