• Take an appropriate wipe or moisten the cloth with cleaner and wipe
the
image plate surface
softly and evenly.
• Leave the
image plate surface
to dry for about 10 minutes to allow
the solvents to evaporate.
• Use only a dry cloth or blow away possible dust with compressed air
(canned air), to clean the
tray
thoroughly. (DO NOT USE A SCREEN
CLEANER TO CLEAN THE TRAY.)
• Once the image plate surface is dry (after about 10 minutes), check
once again for particles of material and other impurities before placing
the tray and image plate set into the cassette.
7.
Put the tray with the image plate back into the cassette.
Verify that the white phosphor side is oriented to the black tube side of the
cassette and that the shutter does not scratch the image plate.
1.
White phosphor side
2.
Black tube side of the cassette
Warning
: Make sure to slide in the image plate carefully.
Do not let the plate fall down vertically into the cassette!
This causes damages of the phosphor.
8.
When the black tray with the image plate is completely inside the cassette,
close the shutter.
To close the shutter easily, insert a key or a pen into the cassette lock while
closing the shutter. Once closed, remove the key or the pen.
Cleaning the image plate and tray of CR MD4.0T and CR MM3.0T | Cleaning the Tray and Image
Plate |
7
0043A EN 20130329 1509