DR Detector is Exceeding the Maximum Working
Temperature
Details
A message is displayed on NX indicating that the DR De-
tector is exceeding the maximum working temperature.
Cause
Due to ambient temperature conditions and the number
of acquired images, the DR Detector’s internal tempera-
ture may become too high.
Brief Solution
1.
Power off the DR Detector.
2.
Leave the DR Detector unpowered for at least one
hour.
3.
Stop the NX workstation.
4.
Power on the DR Detector.
5.
Start the NX workstation.
DR Detector must be Recalibrated
Details
A message is displayed on NX indicating that the DR De-
tector must be recalibrated.
Cause
A DR Detector must be recalibrated at regular interval.
Brief Solution
Follow the instructions in the DR System Key User Man-
ual to calibrate the DR Detector:
• DX-D DR Detector Calibration Key User Manual,
document 0134
DR Detector Problem
Details
An error message is displayed on NX indicating a prob-
lem related to the DR detector.
Cause
-
Brief Solution
1.
Power off the DR detector.
2.
Stop the NX workstation.
3.
Power on the DR detector.
4.
Start the NX workstation.
70
| DX-D 300 | Problem solving
0172E EN 20181116 1653