Image formats
DICOM (Default)
TIFF
Postscript
Not available
Power consumption - heat dissipation
During operation
250 W - 900 kJ/h
In stand-by
70 W - 252 kJ/h
Peak power (absolute max. rating)
530 W - 1908 kJ/h
Protection against
Electrical shocks
Class 1 (grounded)
Ingress of water
IPXØ
Environmental conditions (operation)
Room temperature
B15°C and +35°C
Relative humidity
Between 20% and 75%
Note: Films may not become wet!
Atmospheric pressure
70 kPa - 106 kPa
Related altitude on site
3000 m to 0 m
Environmental storage conditions
Climate conditions for storage are in accordance with EN60721-3-1-class 1K4.
Room temperature
Between -25°C and 55°C (storage)
Relative humidity
Between 10% and 100%
Absolute humidity
Between 0.1 g/m
3
and 35 g/m
3
Rate of change of temperature
1°C/min
Atmospheric pressure
70 kPa - 106 kPa
Environmental transport conditions
Climate conditions for transport are in accordance with
EN60721-3-2-class
2K4.
Temperature
Between -40°C and 70°C (transport)