4.
Compare the measured A and B dimensions with the reference dimension
values, A
ref
and B
ref
on Chart 4 (‘Geometric Consistency Control Chart’).
The differences between measured dimensions of A and B and the
reference values A
ref
and B
ref
should be less than or equal to 1.0%.
5.
Check for image distortion.
6.
Calculate the aspect ratio by dividing A by B.
The result must be 1 +/- 0.01
WARNING:
If the image size or distortion values are outside of limits,
contact your local service organization to resolve the problem.