60
AgaMatrix Jazz Wireless 2
System Specifications
15
CHAP
T
E
R
TECHNICAL SPECIFICATIONS
Assay Method:
Dynamic Electrochemistry
Maximum Altitude:
10,000 feet
Calibration:
Plasma equivalent
Coding:
No Code
Control Solution Storage Temperature:
36°F to 86°F
(2°C to 30°C)
Sample:
Whole blood, capillary
Blood Sample Size:
0.5 microliters
Average Glucose Test Time:
5 seconds
Measurement Units:
mg/dL
Result Range:
20 to 600 mg/dL
Hematocrit:
20% to 60%
Operating Relative Humidity Range:
10% to 90%
System Operating Temperature Range:
50°F to 104°F
(10°C to 40°C)
batteries. Sync meter with your supported device before
attempting another test. If this problem continues please call
customer service.
Troubleshooting Situation #8:
The time and date on the me-
ter do not match the time and date on the supported device.
CAUSE:
Meter has not recently synced with your supported
device.
ACTION:
See Chapter 5 to sync your meter with your
supported device.