Menu “Measure”
MAXILYZER NG Plus
39
Extra menu “Graphic”
This functions uses graphs to show the numerical values according
to the chosen fuel. The remaining content of oxygen (O
2
) and the
calculated waste gas losses (qA) are thereby set in a relation to the
excess air value (
λ
) and to the classical combustion diagram.
If both bars extend to the optimal fuel-air relation (the gap indicated
by “
λ
opt”) the firing facility in question is set in the correct way.
Open the “Graphic” extra menu.
Key
Function
Change representation of readings in the main menu line
by line (multi-tasking-function).
Close menu.