Series 300, 305, 500 & 505 

Monitor 

 

User Guide 

200 

Summary of Contents for 200 series

Page 1: ...1 Series 300 305 500 505 Monitor User Guide 200 ...

Page 2: ...he monitor to gas concentrations above the recommended maximum range Do not place the sensor head in a gas stream the monitor is designed to measure ambient levels of ozone and uses active sampling to achieve this 1 Sensor Head 2 Power and Stand By 3 Display 4 Enter 5 Battery case For battery installation see section 3 0 of your user guide ZERO CAL Please refer to section 4 5 of your user guide wh...

Page 3: ... Key Functions 11 4 5 Zero Calibration 11 4 6 Changing Unit of Measurement 12 4 7 Max Min Average Measurement Cycle 12 4 8 Alarm Mute Perc only 13 4 9 Alarm Perc only 13 4 10 Alarm Level Perc only 14 5 0 Troubleshooting 15 6 0 Specifications 17 7 0 Drawings 18 8 0 Care and Maintenance 18 9 0 For Your Safety 20 10 0 Statements of Compliance 21 11 0 Warranty 22 12 0 Copyright 24 13 0 Guidelines on h...

Page 4: ...ual by email at sales aeroqual com if any of the components are missing Replacements Replacement batteries sensor heads and other accessories can be purchased separately Note All sensors deteriorate over time It will be necessary to replace your sensor head with either a new or refurbished head This must be done on a scheduled basis or when the sensor aging or sensor failure diagnostic appears Bum...

Page 5: ...d This affords the user the ability to replace faulty sensor heads without having to replace the monitor The monitor can be mains and or battery pow ered and either permanently mounted or used as a portable unit 2 The Aeroqual Series 200 Monitor comes with in built diagnostics which will inform the user if the sensor is not operating correctly 3 An audible alarm is available with perchloroethylene...

Page 6: ...the batteries must be installed and charged prior to use 3 1 Battery Pack Remove the battery cover from the monitor Insert the battery power connector into the monitor power socket in the back of the monitor Lay the battery cable into the slot on the back of the monitor and insert the tail end of the battery pack into the hinge points at the bottom of the monitor back cover Clip the top end of the...

Page 7: ...ng the Aeroqual monitor Do not disassemble or mutilate the batteries as this may cause burns Do not incinerate or heat as this may cause burns the batter ies may burst or cause the release of toxic materials Do not short circuit as this may cause burns Use only batteries approved by Aeroqual and recharge your battery only with the AC DC adaptors supplied by Aeroqual Never use any charger or batter...

Page 8: ...ed or removed from the base Remember to switch off the monitor before removing the sensor head Unplugging the main power will reset the above charging condi tions The batteries can be charged and discharged many times but will eventually wear out When the operating time is noticeably shorter than normal it is time to replace the batteries It is not necessary to fully discharge the batteries When n...

Page 9: ... message sensor warming up or sensor failure in older models to disappear Should the message remain after this time a genuine sensor failure may have occurred See the trouble shooting guide Since O3 is an oxidising gas the warm up period can be reduced by subjecting the sensor to elevated O3 levels for a short period For example 100 ppb of O3 for 5 minutes typically reduces the warm up time to 30 ...

Page 10: ... x After 5 seconds the display will change to indicate the type of sensor head The monitor will warm up as described in Warm up in section 3 2 Once warm up is complete the main display will appear for example To turn the monitor off press and hold the power button for 2 seconds until the monitor beeps The display will now be blank and operation of the monitor will cease Power button Enter Key O3 L...

Page 11: ...he sensor runs at full power and the fan remains on 4 2 Operating under Battery power A battery indicator will appear at the bottom of the display when the monitor is operating under battery power only The indicator shows the battery life remaining 4 3 Sensor Head Failure There are two possible Sensor failure modes which will be indicated by one of the following messages on the display Sensor Fail...

Page 12: ...erature at 20 C 2 C 68 F 3 5 F Relative Humidity of 50 5 An incorrect ZERO CAL does not cause permanent damage The process should be repeated under the above conditions NOTE Carbon Dioxide sensors MUST be calibrated in a cer tified Zero Air environment If in doubt contact Aeroqual Limited Press and hold until the word GO appears next to Reset Routine The routine will run for up to ten minutes depe...

Page 13: ...cle Pressing starts the Min Max Av routine Press to scroll to EXIT Press to Exit out of Menu Setup The following display will appear Repeat the above sequence to stop the Min Max Av routine Note The MIN MAX and AVE readings are the minimum maximum and average readings over the period from the start of the cycle The ST value is the average over the last 15 minutes OZONE UNIT ppm or mg m3 O3 LOW M I...

Page 14: ...M MUTE OFF and ALARM MUTE ON Press to confirm your choice and you will return to the main menu If you select Alarm Mute On and exit the menu the symbol will be displayed on the gas concentration screen Press to scroll down to the next menu choice 4 9 ALARM for PERC only To enter the ALARM menu press and the following screen will appear Press to toggle between ALARM PERMANENT ON and ALARM PERMANENT...

Page 15: ...the follow ing screen will appear Press to toggle between 50PPM 100PPM 150PPM and 35PPM Press to confirm your choice and you will return to the main menu Whatever alarm level you have chosen will be displayed on the bottom left of the main display screen NOTE The altered alarm level will remain active until the Moni tor is switched off after which it will revert to the 50PPM default level Press to...

Page 16: ...ensor Reading high under zero gas conditions Background gas level higher than normal Interferent gas present Sensor damaged Move sensor to clean air and recheck baseline Move sensor to clean air and recheck baseline Replace sensor Reading lower than expected reading in the presence of sen sor gas Sensor correct Sensor inlet contaminated Sensor fan failed Interferent gas present Gas reactive and de...

Page 17: ...bration of gas generator Move sensor to clean air and check reading upon exposure to known gas concentration Replace refurbish sensor Reading unstable Power supply unsta ble Power supply current rating incorrect Local air flow too high Environmental condi tions fluctuating Install stable power supply Install power supply with correct rating Reduce air flow Reduce fluctuations ...

Page 18: ...l signal for alarms control None Series 200 Monitor Base External signal functions None Analogue output None Power supply 12 VDC 800 mA Rechargeable battery pack 9 6V 2100mA hr Ni MH Portable Yes Permanently fixable Screw fix Remote Sensor Capability Yes Enclosure rating IP20 NEMA 1 equivalent Size with sensor head 195 x 122 x 54 mm 7R x 4 x 2T in Weight with sensor head and battery 460 g 16 oz Ap...

Page 19: ...18 Depressions in the battery cover demarcating the screw hole posi tions for fixing the unit to the wall 7 0 Drawings ...

Page 20: ...sation that may damage the electronic componentry Do not attempt to open Non expert handling of the device may cause damage Do not drop knock or shake as this could lead to internal damage Do not use harsh chemicals cleaning solvents or strong detergents for cleaning Wipe with a soft cloth slightly dampened with a mild soap and water solution In order to maintain measurement accuracy Aeroqual reco...

Page 21: ...TO OTHER DEVICES When connecting to any other device read the appropriate user guide for detailed safety instructions Do not connect in compatible products HAZARDOUS ENVIRONMENTS Ozone Sensors Only Do not use the Ozone Sensors in or near volatile fuel or chemi cals HEALTH AND SAFETY IN THE WORKPLACE The Aeroqual Series 200 Monitors and Sensor Heads are used to monitor ambient gas concentrations Ae...

Page 22: ...rovide reasonable protection against harmful interference in a residential installation This equipment gener ates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful inter ference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause...

Page 23: ...consumer and for a subsequent period as stated in the following table This warranty is expressly limited to the original owner who purchases the equip ment directly from Aeroqual or from an authorized Aeroqual dealer What we will do If during the warranty period this product fails to operate under normal use and service due to improper materials or workmanship Aeroqual subsidiaries au thorized dis...

Page 24: ...ch have been caused by use of non Aeroqual original accessories This warranty becomes void if a non Aeroqual approved AC DC adaptor or battery is used Tampering with any part of the product will void the warranty Damage to the sensors can occur through exposure to certain sensor poisons such as silicones tetraethyl lead paints and adhesives Use of Aeroqual sensors in these environments containing ...

Page 25: ... the prod ucts described in this document without prior notice Under no circumstances shall Aeroqual be responsible for any loss of data or income or any special incidental consequential or indirect dam ages howsoever caused The contents of this document are provided as is Except as required by applicable law no warranties of any kind either express or implied including but not limited to the impl...

Page 26: ...qual Ozone Controller has been designed to meas ure the ambient concentration of ozone The controller must not be placed directly in an ozone stream For indoor local area monitoring attach the controller to an inert surface with the inlet unobstructed For leak detection mount the unit near the ozone equipment Ensure that the controller is protected from excessive water splashing dust vibration exc...

Page 27: ...r has not been used in the past week There are trace levels of gases present in the air to which the sensor is cross sensitive In many cases the non zero base line reading occurs in an environment which is supposed to be clean but for which no analysis has been carried out such as an office or urban outdoor air Re check the base line reading in clean air from a known source such as a zero air cyli...

Page 28: ...bon NMHC Isobutylene standard 0 25 ppm Ozone O3 0 0 150 ppm 0 0 5 ppm 0 10 ppm Leak detector Perchloroethylene C2Cl4 0 200 ppm Sulphur dioxide SO2 0 10 ppm 0 100 ppm Volatile Organic Compounds VOC Isobutylene standard 0 25 ppm 0 500 ppm Other gases and calibrations available on request contact Aeroqual 15 0 Sensors and accessories Sensors heads are interchangeable R40 Carry Case R33 Wall Bracket R...