EDACS REPEATERS
2-22
AF connections
For tests which require parametric measurements to be made, connections must be made from the
AF GEN OUTPUT connector of the Service Monitor to the audio input of the repeater and from
the audio output of the repeater to the AF INPUT of the Service Monitor. See Fig. 2-10.
Fig. 2-10 A typical test set-up showing connections to a Repeater
Automatic testing
With a repeater connected to the Service Monitor and with the 'SET-UP' configured to suit the
repeater, an
automatic test
can be carried out.
Select the program to be run (e.g. GENERAL TESTING) by using the
[PROGRAM]
,
[general]
,
[AUTO]
, sequence. The selected program is shown at the top of the display.
The results of the test program are displayed and placed into the results store. To make a
permanent record of the results when the program has finished, you can print out a hard copy
using the
[print store]
key.
To run the program press the
[start]
key.
When an AUTORUN program is running , soft keys
[stop]
and
[pause]
are displayed.
Pressing the
[stop]
key stops the program immediately and aborts the current test. The STATUS
message *FAIL* ABORTED BY USER is displayed and a
[cont]
key appears. Pressing this
continues the AUTORUN program, commencing at the next test. Pressing the
[stop]
key rather
than the
[cont]
key prevents the program continuing and prepares for the AUTORUN program to
be started again.
Pressing the
[pause]
key while a program is running makes the program pause at the end of the
current test. The program is then in the same condition as when paused due to your setting of the
PAUSE MODE in the AUTORUN CONTROL menu.