Event: Memory Filled During Recording Session
If the logger is recording using the Start/Stop mode and memory is filled
before the Recording Session has finished, the session will end.
The following happens:
The RED LED double-blinks
•
The GREEN LED single-blinks
•
At this time:
• The memory can be downloaded and cleared
• A new recording can be started or scheduled once the memory is
cleared
Event: Battery Power is Insufficient for a Full Recording Duration
If the battery voltage drops below 1.7V, the following will occur:
The Recording Session will terminate
•
The Data will be saved
•
The GREEN and RED LED will turn OFF
•
The logger continues to record until the battery voltage drops below 1.7V.
Pressing the
PRESS
button may not turn the unit ON at all. The battery
voltage may rise slightly after the unit turns itself off. In this event, the unit
may turn on momentarily as a result of a button press.
The batteries must be replaced before the recorded session(s) can be
downloaded from the instrument.
NOTE:
Replacing the batteries while the unit is OFF will not result in the loss
of data memory. The internal backup capacitor will maintain the clock and
memory while the main batteries are being replaced. If the battery level falls
below the usable level or if the batteries have been removed for an extended
period of time, the clock time will be lost. However, the recorded memory will
be maintained since it is stored in Flash memory.