9
PARAMETER
FUNCTION
Date / Time
Sets the date and time format
Display
Adjusts the contrast and brightness of the display
Defines the color of the voltage and current curves
Calculation Methods
Determines whether or not harmonics are used in calculations of
reactive quantities (power and energy)
With harmonics:
Harmonics are taken into account when
calculating reactive parameters.
Without harmonics:
Only the fundamental part is used for the
calculation of reactive parameters
Electrical Connection
Determines the type of connection to the network
Single-Phase
Split-Phase
3-Phase 4-Wire
3-Phase 5-Wire
Sensors & Ratios
Defines the type of current probe to connect
MN93: 200A
AC
MN193: 100A or 5A
AC
(with variable ratio)
SR193: 1000A
AC
J93: 3500A
AC
/5000A
DC
AmpFlex
®
Sensors: 6500A
AC
MiniFlex
®
Sensors: 6500A
AC
MR193: 1000A/1200A
AC/DC
SL261: 100A (sensitivity
10mV/A
AC/DC
)
10A (sensitivity 100mV/A
AC/DC
)
5A three-phase adapter (3-channel only)
Transient Mode
Configures the voltage and current thresholds
Trend (Recording)
Mode
Selects the parameters to record (up to 4 configurations)
Alarm Mode
Defines the parameters of an alarm
Erase Memory
Deletes configurations, alarm settings, snapshots and recordings
About
Displays the serial number, software and hardware version