AC/DC Current Probe Models MR461 and MR561
13
Frequency Range:
5 to 15Hz, Amplitude: 1.5mm
15 to 25Hz, Amplitude: 1mm
25 to 55Hz, Amplitude: 0.25mm
Handle:
UL 94 V0
Jaws:
UL 94 V0
Color:
Dark gray with red jaws
Output:
Double insulated 6.5 ft (2m) coaxial cable with BNC connector
3.4.3 SAFETY
Electrical:
Double insulation or reinforced insulation between primary, secondary and
outer case of handle per IEC 1010-2-032.
600V, Category III, Pollution Degree 2
300V, Category IV, Pollution Degree 2
7850V
DC
dielectric between primary or secondary and outer hand-held
case
Electromagnetic Compatibility:
• Susceptibility in accordance with EN 50082-2 and EN 50082-1
• Electrostatic discharge IEC 1000-4-2
Test voltages: 4kV level 2 in contact, aptitude criteria B
8kV level 3 in the air, aptitude criteria B
• Radiated Field IEC 1000-4-3 (1995) with max. interference of 5% of
the measurement range: 3V/m level 3, aptitude criteria A
• Fast transients IEC 1000-4-4 (1995)
Test voltage: 1kV level 2, aptitude criteria B
• Magnetic fields at the frequency of the network to IEC 1000-4-8 (1995)
with a max. distortion of 0.5A: 30A/m 50 Hz level 4, aptitude criteria A
• Emissions in accordance with EN 50081-1
• Radiated emission through the case to EN 55022 (1994): class B
• Conducted emission to EN 55022 (1994): class B
*Specifications are subject to change without notice
www.
ShopAEMC
.ca
Shop for AEMC products online at:
1.800.561.8187