OPERATION

 

Please make sure that you have already read and fully understand the WARNING section on page 1. 

Making Measurements with the AC Current Probe Model MN251T:

 

•  Connect the BNC connector to your voltage-measuring instrument. The AC current probe has an output of 

1mV/A. This means that for 200A AC in a conductor around which the probe is clamped, 200 mV AC will 
come out of the probe leads to your DMM or instrument. Select the range which corresponds to the 
measured current. If the current magnitude is unknown, start with the highest range and work down until 
the appropriate range and resolution are reached. Clamp the probe around the conductor. Take the 
reading on the meter and multiply it by 1000 to obtain the measured current (e.g, 150mV reading = 150 x 
1000 = 150,000mA or 150A). 

•  For best accuracy, avoid if possible, the proximity of other conductors which may create noise. 

Tips for Making Precise Measurements:

 

•  When using a current probe with a meter, it is important to select the range that provides the best 

resolution. Failure to do this may result in measurement errors. 

•  Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air 

gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for 
power measurement. 

 

MAINTENANCE

 

Warning:

 

•  For maintenance use only original factory replacement parts. 

•  To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so. 

•  To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into 

the probe. 

Cleaning:

 

To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times. 
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine 600) 
to avoid scratching the jaw, then gently clean with a soft oiled cloth. 

 

REPAIR AND CALIBRATION

 

You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure 
that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the 
outside of the shipping container. 

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

® 

Instruments 

(Or contact your authorized distributor) 

NOTE: All customers must obtain a CSA# before returning any instrument.

 

 

TECHNICAL AND SALES ASSISTANCE

 

If you are experiencing any technical problems, or require any assistance with the proper use or application 
of this instrument, please contact our technical hotline: 

LIMITED WARRANTY

 

The current probe is warranted to the owner for a period of two years from the date of original purchase 
against defects in manufacture. This limited warranty is given by AEMC

®

 Instruments, not by the distributor 

from whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the defect 
is related to service not performed by AEMC

®

 Instruments. 

Please print the online Warranty Coverage Information for your records. 

 

 