16
DC/AC MicroProbe Models K100 and K110
4.3.3 Two-Wire Sum-of-Currents Example
(K100)
• Two conductors in probe aperture (note orientations)
• Voltmeter placed in DC Volts mode
• Voltmeter displays 18.0 mV
PUSH
DC Zero
18.0
mV
AC
V
DC
V
7 mA DC
11 mA DC
Figure 7
4.3.4 Two-Wire Current Differential Example
(K100)
• Two conductors in probe aperture (note orientations)
• Voltmeter placed in DC Volts mode
• Voltmeter displays -4.0 mV
PUSH
DC Zero
-4.0
mV
AC
V
DC
V
7 mA DC
11 mA DC
Figure 8