DC/AC MicroProbe Models K100 and K110
19
4.4.3 Oscilloscope Measurement Example
(K110)
• Conductor carrying a 0.2 A peak AC waveform
• Model K110 connected to oscilloscope
0
To Oscilloscope
Current Waveform
Oscilloscope
Output
• OFF
• 10 mV/mA
ON
OL
AEMC
I N S T R U M E N T S
®
DC/AC MicroProbe
Model K110
0
0.2 A
2.0 V
-2.0 V
-0.2 A
Figure 13
4.5 TIPS FOR MAKING PRECISE MEASUREMENTS
The Model K100/K110 is capable of making measurements of DC and low
frequency currents over a wide range. Here are some key considerations for
getting the most accuracy from your displaying instrument:
• When using the Model K100/K110 with a meter, it is important to select
the range that provides the best resolution. Failure to do this may
result in measurement errors. The examples shown on the next page
use an ordinary 3-1/2 digit DMM with the Models K100 and K110, to
measure 16.7 mA
DC
.