AC/DC Current Probe Model J93
7
4. SPECIFICATIONS
4.1 Reference Conditions
Quantity of influence
Reference Values
Temperature
23°C ±5°C
Relative humidity
20 to 75% RH
DC signal
With an AC signal of which the distortion factor is <0.1%
AC signal
45 to 65Hz with a DC signal <0.1%
Supply voltage
J93: 9V ± 0.1V
External electric field
Zero
External DC magnetic field
(earth’s field)
< 40 A/m
External AC magnetic field
Zero
Position of the conductor
centered in the probe jaws
4.2 Electrical
Measurement Range
50A to 3500A
ac
50A to 5000A
dc
Sensitivity
1V / 3500A
Output impedance
≥
100k
W
Frequency
DC to 3kHz (-3 dB typical)
Primary
Current
(AC/DC)
50 to 100
100 to 500
500 to 2000 2000 to 3500 3500 to 5000
(DC
only)
Amplitude
error
(see curve
in §4.4.1)
± 2% ± 2.5A ± 1.5% ± 2.5A
± 1%
± 1%
± 1%
Phase error
(see curve
in §4.4.2)
4°
2°
1°
1.5°
‒
Zero adjustment: ± 200A in 25 turns