12
Simple Logger
®
II Model CL601
4.2
Connecting the Simple Logger
®
II to a Computer
INSTALL DATAVIEW
®
BEFORE CONNECTING TO THE COMPUTER.
Connect the Simple Logger
®
II to a USB communication port on your
computer. Refer to the computer manual to locate the USB port on your
computer.
The logger can be connected to the computer during a recording session,
however, additional battery drainage will occur to support the active USB
connection.
4.3
Turning the Unit On
Turn the instrument ON by performing one of the following:
• Press and hold the
PRESS
button for approximately 2 seconds.
Release the button after all five LEDs light up in unison.
The unit is now in the
STANDBY
mode (Green LED single-blinks).
NOTE:
If all LEDs light instantaneously, the instrument was in
SLEEP
mode.
Releasing the
PRESS
button will show the status.
• Connect the instrument to a USB port on your computer and establish
communication with the instrument using DataView
®
. The logger will
remain ON while a communication link with the SLII Control Panel is
active (provided sufficient battery power is available during the com
-
munication session).
The instrument contains protection circuitry to prevent it from being turned
on when the battery voltage is below 1.7V.
There are two thresholds for the battery voltage:
•
The first is used to indicate low battery.
The low battery indicator will
blink when the battery voltage drops below 1.8V.
•
The second is used to determine when to terminate recording and
turn the unit off.
The shutdown threshold is when the battery voltage
drops below 1.7V.