Fault Mapper Pro

®

 Model CA7027

 

11

 

NOTE: 

The auto-off function is disabled in Tone Generator mode, so that 

the tone can be injected into a cable for an extended period of time while 

tracing takes place.

See §4.10 for attaching a cable to the Fault Mapper Pro

®

4.7 

Selecting Full Scale Sensitivity

The Fault Mapper Pro

®

 has 11 range scales covering the range of 1.5 to 

19,000 ft (0.5 to 6000m).

To select a range scale, or scan the cable run, press and hold down the 

TDR /

  button, then press the   button to decrease the range or the     

button to increase the range.

4.8 

Adjusting the Vp Value

To adjust the Vp, press and hold the 

TDR /

 button and press the 

TEST /

 

button to enter the setup mode. Release both buttons. Press the 

TEST /

 

button until the pointer (

>

) is next to Vp.

Next use the 

 and 

 buttons to adjust the Vp to scroll between v/2, M/Ms 

or Vp % UC.

Press the 

TDR /

 button to exit the set up mode.

4.9 

Determining and Measuring Vp Values

Vp, or Velocity of Propagation, values are characteristic of each cable 

type and brand. The Vp is used to measure the length of a cable and to 

measure a fault location. The more accurate the Vp, the more accurate the 

measurement result will be.
The cable manufacturer may list the Vp on their specification sheet or may 

be able to provide it when asked. Sometimes this value is not readily avail

-

able, or the user may wish to determine it specifically to compensate for 

cable batch variations or for special cable applications.
1.  Take a cable sample of exact length increments (ft or m) longer than 

300 ft (100m).

2.  Measure the exact length of the cable using a tape measure.

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for CA7027

Page 1: ...CA7027 FAULT MAPPER PRO GRAPHICAL TDR E N G L I S H User Manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...et its published specifications The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer Serial _________________________________ Catalog...

Page 3: ...of Propagation Vp 8 4 3 Getting Started 9 4 4 Set up Mode 9 4 5 Backlight 10 4 6 Tone Generator 10 4 7 Selecting Full Scale Sensitivity 11 4 8 Adjusting the Vp Value 11 4 9 Determining and Measuring V...

Page 4: ...strument is protected by double or reinforced insulation CAUTION Risk of Danger Indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument...

Page 5: ...on of any damage Save the damaged packing container to substantiate your claim 1 3 Ordering Information Fault Mapper Pro Model CA7027 Cat 2127 84 Includes carrying case 4mm color coded banana plugs wi...

Page 6: ...f the speed of light e g 67 or ft mtrs micro second this value will vary according to the type of cable under test The Fault Mapper Pro can accept user selectable values between 1 and 99 or the equiva...

Page 7: ...1 1 Banana plug input Red Black 2 Graphical LCD 3 Cable select and Vp Velocity of Propagation decrement button 4 Test function select button 5 Backlight button 6 Cable select and Vp Velocity of Propa...

Page 8: ...scanning Auto off Selectable 1 2 3 5 minutes or disabled MECHANICAL Dimensions 6 5 x 3 5 x 1 5 165 x 90 x 37mm Weight with battery 12oz 350g Display Resolution 128 x 64 pixel Graphical LCD Display Bac...

Page 9: ...he Vp along the cable length If the Vp is set incorrectly by the operator or the Vp varies along the length of the cable then additional errors will be incurred and the measurement accuracy will be af...

Page 10: ...s based upon the cor rect value of Vp being used for the cable under test and homogeneity of the Vp along the cable length If the Vp is set incorrectly by the operator or the Vp varies along the lengt...

Page 11: ...C Vp 67 T P 100 Meters Contrast Shutdown 1 Minute 2 The following options can be configured TDR or Tone Generation Set Vp to 1 to 99 or speed in micro seconds see 4 8 Set cable impedance Z to 25 50 75...

Page 12: ...nditions The backlight is switched on and off with the button 4 6 Tone Generator The Fault Mapper Pro may also be used as a tone generator to trace and identify cables and wires The user will need a c...

Page 13: ...ter is next to Vp Next use the and buttons to adjust the Vp to scroll between v 2 M Ms or Vp UC Press the TDR button to exit the set up mode 4 9 Determining and Measuring Vp Values Vp or Velocity of P...

Page 14: ...ve termination 2 Attach the Fault Mapper Pro to one end of the cable to be tested The cable attachment is via a banana plug input located at the top of the unit For unterminated cables use the alligat...

Page 15: ...right along the line of the trace by pressing the and buttons To determine the distance to an event position the cursor at the beginning of the event and read off the distance as shown below Vp 72 Z...

Page 16: ...he bottom right of the display when Continuous Scanning mode is active Continuous Scanning Mode Fires pulses into the cable under test at the rate of 6 7 pulses per second In this mode the Fault Mappe...

Page 17: ...cover 3 Replace the batteries with 4 x 1 5V AA alkaline batteries observing the polarities 4 Reattach the battery compartment cover 5 2 Cleaning Disconnect the instrument from any source of electrici...

Page 18: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 19: ...us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material Warranty Repairs What yo...

Page 20: ...01 12 99 MAN 100361 v2 Chauvin Arnoux Inc d b a AEMC Instruments www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Reviews: