Fault Mapper Pro
®
Model CA7027
9
4.3
Getting Started
The instrument is switched on and off using the green power button
,
found on the lower right side of the front panel. When the unit is first switched
on it will display the opening screen giving the software version, the currently
selected Vp (Velocity of Propagation), cable impedance and battery status
indicator.
AEMC Instruments
Fault Mapper Pro
VX.X
4.4
Set-up Mode
1.
Hold down the
TDR /
button, then press the
TEST /
button to enter
set-up mode.
TDR
Vp % VC
Vp=67%
T/P 100
Ω
Meters
Contrast
Shutdown : 1 Minute
>
2.
The following options can be configured:
• TDR or Tone Generation
•
Set Vp to % (1 to 99%) or speed in micro-seconds (see § 4.8)
• Set cable impedance (Z) to 25, 50, 75 or 100
Ω
Twisted pair (T/P)
• Set measurement units to Feet or Meters
• Adjust the display contrast
• Disable/Enable Auto-off function
3. Press the
TEST /
button to move the line selector (
>
) down the screen.
4. Press the and buttons to change the setting of the line selected.
5. Press the
TDR /
button again to save changes and exit set-up mode.
NOTE: When the Fault Mapper Pro
®
is turned off, it will retain the
current set-up parameters.