ITEM

BUTTON

DESCRIPTION

9

SENSITIVITy BUTTONS

 

-

decreases

 the vertical sensitivity of the last selected channel down to 5mV.

 

-

increases

 the vertical sensitivity of the last selected channel up to 200V.

NOTE: 

For the M channel, the “sensitivity” buttons vary the amplitude factor 

but only if a math channel is validated.

10

MENU BUTTONS

 

- displays the main 

“Trigger”

 menu

 

- displays the main 

“Tools” 

menu

 

- displays the main 

“Acquisition” 

menu

 

- displays the main 

“Memory” 

menu

 

- displays the main 

“Measurement/Cursor”

 menu

 

- displays the 

“Help”

 window

Use of the Sensors

Distribution of stray capacitors:

Considering the stray capacitances, it is imperative to correctly connect the reference conductors for 
each sensor. The conductors should preferably be connected to the cold points to avoid the transmis-
sion of noise by the stray capacitance between modes.

             

The noise of the digital ground (earth) is sent to the analog input by the stray capacitance.

  

NOTE: 

In order to prevent electric shocks or possible fires, never use accessories on which the 

casing is accessible if it has a voltage of > 30Vrms compared to the ground.

www.

ShopAEMC

.com

Shop for AEMC products online at:

1.877.766.5412

Summary of Contents for 2150.20

Page 1: ...QUICK START USER GUIDE OX 5022 CK OX 5042 CK www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...

Page 2: ...ime of purchase or obtained by returning the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on...

Page 3: ...0 20 2150 21 Also Included 1 Test Report 1 Battery Information Sheet 6 1 2V NiMH rechargeable batteries 2700mA h 1 Soft Hand held Carrying Case Cat 5000 48 2 CD ROMs software and manual 1 USB Cable Dr...

Page 4: ...nsors or test probes do not place your fingers beyond the physical guard If the battery housing cover is absent damaged or incorrectly positioned the instrument must not be used other than to adjust t...

Page 5: ...between 40 and 450Hz The instrument operates at a constant acquisition depth of 2500 points An LCD TFT screen is used to view the applied signals along with all the setting parameters The main command...

Page 6: ...Control Features Measurement Terminals 1 2 3 4 5 10 9 8 7 6 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...

Page 7: ...math memory buttons A single press selects channel M Math and shows the corresponding menu Pressing twice deselects the channel NOTE For the M memory channel pressing twice invalidates the channel Pr...

Page 8: ...rement Cursor menu displays the Help window Use of the Sensors Distribution of stray capacitors Considering the stray capacitances it is imperative to correctly connect the reference conductors for ea...

Page 9: ...djustment screw In this example the sensor is over compensated an overshoot occurs Turn the screw in either direction until the signal is horizontal and looks like the screen shown opposite Your senso...

Page 10: ...www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...

Page 11: ...nty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material Warranty Repairs What you must do to return an Instrument for Warr...

Page 12: ...10 13 99 MAN 100382 v2 Chauvin Arnoux Inc d b a AEMC Instruments 15 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...

Reviews: