Power Supply

The oscilloscope is supplied with one external power supply and set of 6 x 1.2V NiMH rechargeable 
batteries 2700mA/h.
When the supplied power cord is connected from the instrument to an external power source, battery 
power is not needed. The batteries are only used when there is no external power supply available.

Before the first use, start by fully charging the battery. 

 

NOTE: 

The batteries automatically begin recharging when the Oscilloscope is powered off, but 

connected to an external power supply.

When the batteries are recharging, the Battery Charge LED on the front panel will turn on steady. 
The LED will blink in the following situations:

• 

pre-charge of very low batteries

• 

temperature is too low or too high

• 

batteries are damaged

When the charge is complete, the LED will turn off. 
If the batteries need to be replaced, they must be replaced with NiMH rechargeable batteries. Battery 
charge life is guaranteed for same-capacity batteries (in mAh) as those shipped with the oscilloscope.
To remove the battery cover, use a coin to turn the slot on the back of the unit counter-clockwise.

Description

The Handscope Oscilloscope Models OX5022-CK and OX5042-CK combine three instruments into one:

• 

20MHz or 40MHz oscilloscope

• 

Independent 8000-count multimeter with power measurements

• 

Built-in harmonic analyzer out to the 31st harmonic (fundamental between 40 and 450Hz)

The instrument operates at a constant acquisition depth of 2500 points.
An LCD TFT screen is used to view the applied signals along with all the setting parameters.  
The main command functions are accessible using the buttons on the front panel. 
A graphic interface is used to:   

 

-

adjust the parameters related to the selected button

 

-

navigate using a horizontal main menu showing the current settings and vertical sub-menus

Channel Insulation

The two oscilloscope input channels are insulated from each other and from the ground/earth and the 
main power supply block.  This insulation is double or reinforced in compliance with the safety standards 
IEC 61010-1 and IEC 61010-2-030.
This makes it possible to make measurements on installations or systems connected to the electricity 
supply network for voltages of up to 600V in CAT III. The common mode authorized between the two 
channels is 600V in CAT III. Thus the operator, the test systems and the environment are completely 
protected at all times.
Any voltage (even dangerous) on one channel will not be present on the other channel. The low points 
of the inputs are completely insulated, so there is no possibility of the low points looping (which can be 
dangerous and highly destructive).

 

NOTE: 

The use of accessories with a voltage and/or category lower than 600V CAT III reduces 

the operating range to the lower voltages and/or categories. 
The oscilloscope is rated 600V CAT III; at least 600V CAT III accessories must be used.

www.

ShopAEMC

.com

Shop for AEMC products online at:

1.877.766.5412

Summary of Contents for 2150.20

Page 1: ...QUICK START USER GUIDE OX 5022 CK OX 5042 CK www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...

Page 2: ...ime of purchase or obtained by returning the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on...

Page 3: ...0 20 2150 21 Also Included 1 Test Report 1 Battery Information Sheet 6 1 2V NiMH rechargeable batteries 2700mA h 1 Soft Hand held Carrying Case Cat 5000 48 2 CD ROMs software and manual 1 USB Cable Dr...

Page 4: ...nsors or test probes do not place your fingers beyond the physical guard If the battery housing cover is absent damaged or incorrectly positioned the instrument must not be used other than to adjust t...

Page 5: ...between 40 and 450Hz The instrument operates at a constant acquisition depth of 2500 points An LCD TFT screen is used to view the applied signals along with all the setting parameters The main command...

Page 6: ...Control Features Measurement Terminals 1 2 3 4 5 10 9 8 7 6 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...

Page 7: ...math memory buttons A single press selects channel M Math and shows the corresponding menu Pressing twice deselects the channel NOTE For the M memory channel pressing twice invalidates the channel Pr...

Page 8: ...rement Cursor menu displays the Help window Use of the Sensors Distribution of stray capacitors Considering the stray capacitances it is imperative to correctly connect the reference conductors for ea...

Page 9: ...djustment screw In this example the sensor is over compensated an overshoot occurs Turn the screw in either direction until the signal is horizontal and looks like the screen shown opposite Your senso...

Page 10: ...www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...

Page 11: ...nty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material Warranty Repairs What you must do to return an Instrument for Warr...

Page 12: ...10 13 99 MAN 100382 v2 Chauvin Arnoux Inc d b a AEMC Instruments 15 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...

Reviews: