Getting Started

 

NOTE: 

Make sure the PowerPad

®

 is fully charged before use.

Connecting:

•  Start the instrument by pressing the 

 button.

•  Configure the instrument to obtain the required results and type of network.
•  Connect the current leads and sensors to the PowerPad

®

.

•  Connect the ground and/or neutral lead to the network ground and/or neutral (when 

distributed), as well as the corresponding current sensor.

•  Connect the L1 phase lead to the network L1 phase, as well as the corresponding  

current sensor.

•  Repeat the procedure for phases L2, L3 and N.

Disconnecting:

•  Proceed in the reverse order to connecting, always finishing by disconnecting the  

ground and/or neutral (when distributed).

•  Disconnect the leads and press the 

 button to turn the instrument off.

•  Recharge the battery, if necessary.

Installation of the Leads and Current Sensors

Twelve sets of color-coded rings and inserts are supplied with the PowerPad

®

. Use these ID 

markers to identify the leads and input terminals.

•  Detach the appropriate inserts from the color-coded marker and place them in the holes 

provided under the terminals (larger inserts for current terminals, smaller inserts for volt-
age terminals).

•  Clip the rings of the same color to the ends of the lead you will be connecting to the 

terminal.

Summary of Contents for 2136.20

Page 1: ...QUICK START USER GUIDE 8335 PowerPad IMPORTANT WARRANTY NOTE By registering online within 30 days from the date of purchase your warranty will be extended to 3 years...

Page 2: ...ent to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For recalibratio...

Page 3: ...aView User Manual NiMh Battery installed Kits will also include order dependent 4 MN93 BK current probes 4 MN193 BK current probes 4 SR193 BK current probes 4 AmpFlex 193 24 BK 24 current sensors 4 Am...

Page 4: ...input Harmonics Mode Displays the harmonics in percent and value ratios for voltage current and power for each harmonic through the 50th Determines harmonic current produced by non linear loads Analyz...

Page 5: ...voltage inputs 7 USB port 8 Input for external power supply and battery charging 9 Confirm Enter button 10 Navigation buttons 11 Six mode buttons see chart left Connection Terminals 3 1 2 1 Four 4 cu...

Page 6: ...voltage and current curves Calculation Methods Determines if harmonics are used or not used in calculations of reactive quantities power and energy With harmonics Harmonics are taken into account when...

Page 7: ...2 L3 and N Disconnecting Proceed in the reverse order to connecting always finishing by disconnecting the ground and or neutral when distributed Disconnect the leads and press the button to turn the i...

Page 8: ...older view to launch the Dataview setup program NOTE For more information on using DataView refer to the Model 8335 user manual that is supplied on the USB stick Updating Software Firmware To provide...

Page 9: ...certificate plus recorded calibration data Ship To Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 742 2346 or 60...

Page 10: ...he one year period you may return the instrument to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or r...

Page 11: ...NOTES...

Page 12: ...10 13 99 MAN 100392 v2 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com...

Reviews: