32
4.2.2 AC Voltage Measurement
Measurement
range
0.15 to 59.99V
60.0 to 599.9V 600 to 1000V RMS
1400V peak (1)
Specified
measurement
range (2)
0 to 100% of the measurement range
Accuracy
0.15 to 5.99V
± (1% R + 10cts)
6.00 to 59.99V
± (1% R +3cts)
± (1% R +3cts)
Resolution
0.01V
0.1V
1V
Input impedance
10M
Ω
Note (1) Above 1000V (RMS), a repetitive beep indicates that the voltage
being measured is greater than the safety voltage for which the
device is guaranteed. The display indicates "OL".
- Bandwidth in AC = 3 kHz
Note (2) Any value between zero and the min. threshold of the measurement
range (0.15V) is forced to show "----" on the display.
Specific Specifications in MAX/MIN mode
(from 10Hz to 1kHz, and from
0.30V in AC):
•
Accuracy: add 1% R to the values of the table above.
•
Capture of the extreme: approximately 100ms.
4.2.3 DC Current Measurement
Measurement
Range (2)
0.00 to
59.99A
60.0 to
599.9A
600 to 900A (1)
Specified
measurement range
0 to 100% of the measurement range
Accuracy (2)
(zero corrected)
± (1% R + 10cts)
± (1% R +3cts)
Resolution
0.01A
0.1A
1A
Note (1) The display indicates “+OL” above 1800A and “-OL” below -1800A
in REL mode. The "-" and "+" signs are displayed.
Note (2) The residual current at zero depends on the remanence. It can be
corrected by the “DC zero” function of the HOLD button.
Summary of Contents for 203
Page 1: ...203 CLAMP ON METER User Manual ENGLISH ...
Page 2: ...2 ...
Page 42: ...42 NOTES ...
Page 43: ......